The Control Cube servo controller
download- 19
- 0.4 kHz
- 4 kHz
- Very suitable for electro-hydraulic servo fatigue test system (uniaxial/multi-axis/reverse)
Bring fatigue test and fatigue strength test maximum flexibility and test efficiency
For fatigue and durability test, the preferred use electro-hydraulic servo fatigue test system.The performance of the system depends largely on the use of electronic measurement and control system, as well as the related test software.ZwickRoell为此提供了Control Cube伺服控制器和Cubus测试软件。它们非常适合新的试验系统、升级改造现有电液伺服试验系统以及多轴应用。
The system consists of ZwickRoell with CaTs3(to be automatic test and structure dynamic simulation system of consultants) joint development.The latter's long-term experience and expertise to achievement the powerful and efficient products, is able to cope with the increasing demand of customers, at the same time meet the high standards of quality and safety ZwickRoell.
Is of particular concern in the process of development, achieve high flexibility and scalability of the test system.
An overview of the technology
- (1) Ethernet PC interface
- (2) the multi-channel control interface (up to 32 channels)
- (3) the optional default plus or minus 10 v channel monitoring
- (4) digital I/O interface
- (5) the remote control interface
- (6) the hydraulic control interface
- (7) system interface (such as exit system)
- (8) stop system interface
- (9) the LED channel identification indicator light
- Gm (10) AC/DC sensor measuring amplifier
- (11) analog I/O module interface (optional)
describe | The numerical | |
Control frequency | 4096 | Hz |
Maximum sampling frequency | 4096 | Hz |
Expansion slot | 3 | |
PC interface | Ethernet | |
Measuring amplifier | 2 x general digital signal converter, each Control Cube up to 8 x;At 64 kHz, on average, have 19 under the sampling frequency resolution | |
Analog signal input and output | 2 x + 10 V analog signal input, can be used for an external control or an external signal amplifier 2 x + 10 V analog signal output, can be used for synchronous control peripheral equipment;Can also be used to signal monitoring, such as external oscilloscope, etc Expansion option |
|
Digital signal input and output | 2 x optical coupling digital signal input 2 x digital signal output;Expansion option |
|
Power options | 85-264 - V AC, 47-63 Hz / 150 W | |
The size (H x W x D) about | 94 x 233 x 315 | mm |
About weight | 8 | kg |